Author:
Otlet R L,Huxtable George,Evans G V,Humphreys D G,Short T D,Conchie S J
Abstract
The Harwell system for measuring milligram size samples using Brookhaven miniature gas counters is fully operative. It comprises 12 counters of different sizes which operate simultaneously within a single NaI crystal (300mm diameter × 300mm long) acting as an anti-coincidence guard counter. Brief details are given of the construction and commissioning of the system, including counter assembly, shield design, electronics, data capture, data analysis, and chemical processing and filling procedures. The performance of the system and an overall view of the fields of application for which the counters have important applications are discussed.
Publisher
Cambridge University Press (CUP)
Subject
General Earth and Planetary Sciences,Archeology
Reference9 articles.