Gottdang Andreas,Mous Dirk J. W.,Van Der Plicht Johannes
Abstract
Since May 1994, a new-generation accelerator mass spectrometer (AMS) has been fully operational at the Centre for Isotope Research in Groningen, The Netherlands. The fully automated and high-throughput accelerator mass spectrometry (AMS) system, manufactured by High Voltage Engineering Europa (HVEE) is dedicated to radiocarbon analysis. The HVEE 4130 14C AMS is able to analyze up to 3000 samples per year. The system is characterized by simultaneous transport of all three isotopes (12C, 13C, 14C) and 14C analysis with a precision below 0.5 pMC and a daily stability below 0.5 pMC. We present here a system description together with stability and performance measurements.
Publisher
Cambridge University Press (CUP)
Subject
General Earth and Planetary Sciences,Archeology
Reference19 articles.
1. Litherland A. E. and Kilius L. R. 1991 Mass recombinator for accelerator mass spec trometry. US Patent No. 5.013.923.
2. Purser K. H. 1992c Accuracy of AMS isotopic ratio measurements. U.S. Patent No. 5.118.936.
3. AMS beyond 2000;Davis;Nuclear Instruments and Methods in Physics Research,1994