The method of applying the scattering correction in accurate γ ray absorption measurements is discussed, and it is shown that the scattering correction only diminishes, as the solid angle subtended by the chamber at the source is reduced, if the absorber diameter is reduced correspondingly. The method of calculation of the scattering correction in cases in which the absorber is placed very close to the chamber is explained, and the result of the calculation for one particular chamber is shown to be in satisfactory agreement with the experimental values. The scattering correction allows the ionization function of chambers to be determined and an (ionization function)/(wave-length) curve is shown for a steel thick-walled high-pressure ionization chamber.