A comparative evaluation of capnometry versus pulse oximetry during procedural sedation and analgesia on room air

Author:

Sivilotti Marco L.A.,Messenger David W.,Vlymen Janet van,Dungey Paul E.,Murray Heather E.

Abstract

ABSTRACTObjective:Important questions remain regarding how best to monitor patients during procedural sedation and analgesia (PSA). Capnometry can detect hypoventilation and apnea, yet it is rarely used in emergency patients. Even the routine practice of performing preoxygenation in low-risk patients is controversial, as supplementary oxygen can delay the detection of respiratory depression by pulse oximetry. The purpose of this study was to determine whether the capnometer or the pulse oximeter would first detect respiratory events in adults breathing room air.Methods:During a randomized clinical trial comparing fentanyl with low-dose ketamine for PSA with titrated propofol, patients were monitored using pulse oximetry and continuous oral–nasal sampled capnography. Supplemental oxygen was administered only for oxygen desaturation. Sedating physicians identified prespecified respiratory events, including hypoventilation (end-tidal carbon dioxide > 50 mm Hg, rise of 10 mm Hg from baseline or loss of waveform) and oxygen desaturation (pulse oximetry < 92%). These events and their timing were corroborated by memory data retrieved from the monitors.Results:Of 63 patients enrolled, 57% (36) developed brief oxygen desaturation at some point during the sedation. All responded to oxygen, stimulation or interruption of propofol. Measurements of end-tidal carbon dioxide varied substantially between and within patients before study intervention. Hypoventilation (19 patients, 30%) was only weakly associated with oxygen desaturation (crude odds ratio 1.4 [95% confidence interval 0.47 to 4.3]), and preceded oxygen desaturation in none of the 12 patients in whom both events occurred (median lag 1:50 m:ss [interquartile range 0:01 to 3:24 m:ss]).Conclusion:During PSA in adults breathing room air, desaturation detectable by pulse oximeter usually occurs before overt changes in capnometry are identified.

Publisher

Springer Science and Business Media LLC

Subject

Emergency Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3