Coronary compression testing by balloon interrogation during pulmonary valve implantation: room for doubt?

Author:

Morgan Gareth J.,Rodriguez Salvador A.ORCID,Leahy Ryan,Zablah Jenny E.

Abstract

AbstractObjective:To evaluate the reliability of balloon coronary compression testing during percutaneous pulmonary valve implantation.Background:Despite the widespread use of the ‘balloon coronary test’ as the preferable method to rule out the risk of coronary compression, this adverse event has been described after pulmonary valve implantation where coronary balloon test suggested no risk or low risk, calling into question the accuracy of the test.Methods:We performed a retrospective chart review of 84 patients who underwent pulmonary valve implantation between January 2018 and December 2019 and selected 36 patients whose archived imaging was suitable to perform quantitative analysis of the ‘balloon coronary test’. We focused on the spatial disparity between the right ventricular outflow tract position defined by the inflated testing balloon and the eventual implanted valve position, to classify the test as inaccurate or accurate.Results:In total, 36.1% of cases were classified as having an inaccurate coronary balloon test. Among the baseline characteristics, right ventricular outflow tract substrate was identified as a significant predictor of test accuracy. Related to this characteristic, the type of testing balloon used and the size of the eventually implanted valve were found to be associated with test accuracy.Conclusions:Based on our findings, balloon coronary testing is not an accurate method of predicting final valve position with respect to fixed structures in the thorax. This may translate to a high false positive rate for the likelihood of coronary compression in pulmonary valve implantation.

Publisher

Cambridge University Press (CUP)

Subject

Cardiology and Cardiovascular Medicine,General Medicine,Pediatrics, Perinatology and Child Health

Reference22 articles.

1. Percutaneous Pulmonary Valve Implantation

2. Coronary anomalies in tetralogy of Fallot – A meta-analysis

3. Three dimensional rotational angiography for assessment of coronary arteries during melody valve implantation: introducing a technique that may improve outcomes;Pockett;Neth Heart J Mon J Neth Soc Cardiol Neth Heart Found,2017

4. The Sapien valve provides enough grip to be implanted in pulmonary position without a pre-stent

5. Risk of Coronary Artery Compression Among Patients Referred for Transcatheter Pulmonary Valve Implantation

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献