Abstract
AbstractBackground:Catheterization in small children should be performed with the lowest diameter introducer sheaths to prevent permanent vessel damage. The objective of this study is to evaluate the clinical safety and efficacy of the Glidesheath Slender in small children.Methods and results:We present a group of 52 patients (male: n = 36) with median age 118.5 days (min. 3; max. 1302), median weight: 5.3 kg (min. 1.4; max. 14.0), median height: 60.5 cm (min. 39; max. 102), and median body surface area 0.28 m2 (min. 0.12; max. 0.63) in whom percutaneous catheter interventions (n = 55) were performed via a Glidesheath Slender. In 49 children, the intervention was performed from femoral access (artery n = 35; vein n = 14) in 2 from the axillary arterial access and in 1 from the jugular venous access. In all patients, the vessel access was obtained under ultrasound guidance. After the catheterization, the pulse on the peripheral arteries (posterior tibial artery or radial artery) was palpable in all patients, and no signs for vessel dysfunction were present.Conclusion:The Glidesheath Slender effectively reduces the outer sheath diameter for various types of interventions in small children by one French, reducing the risk of vessel complications (stenosis, occlusion). Interventions via Glidesheath Slender in small patients are safe and feasible and extend the transcatheter possibilities in small children with congenital heart diseases.
Publisher
Cambridge University Press (CUP)
Subject
Cardiology and Cardiovascular Medicine,General Medicine,Pediatrics, Perinatology, and Child Health
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献