Abstract
Abstract
Objective:
In this study, we describe our experience utilising Advisor™ High Density (HD) Grid mapping catheter in transcatheter ablation of intraatrial re-entrant and focal atrial tachycardias with or without CHD.
Methods:
Forty-five consecutive patients with intraatrial re-entrant and focal atrial tachycardia who underwent a transcatheter ablation procedure by using Advisor™ HD Grid mapping catheter and high-density mapping system in our hospital from January 2017 to January 2023 were included into the study.
Results:
The mean age of the patients was 14.2 ± 7.3 years (6–32 years), and the mean weight was 48.3 ± 16.2 kg (22–83 kg). Of the total 45 patients, 21 were intraatrial re-entrant tachycardia and 25 were focal atrial tachycardia. Of the 21 re-entrant circuits, 15 were classified as cavotricuspid isthmus-dependent and 5 were non-cavotricuspid isthmus-dependent. In one patient, two re-entrant circuits were identified. A transbaffle ablation was successfully performed from the left atrium in one patient. Of the 25 focal atrial tachycardia, 19 were from right atrium and 6 were from left atrium. A cryoablation was performed in only one patient and radiofrequency ablation in others. The mean procedure time was 180 ± 64 minutes. The mean follow-up period was 69.3 ± 35.3 months. Acute success was 95.5%. Recurrence was noted in two patients (4.4%).
Conclusion:
Advisor™ HD Grid mapping catheter was found to be safe and achieved an acceptable success in transcatheter ablation of patients with intraatrial re-entrant tachycardia and focal atrial tachycardias.
Publisher
Cambridge University Press (CUP)
Subject
Cardiology and Cardiovascular Medicine,General Medicine,Pediatrics, Perinatology and Child Health