Polonenko M J,Cushing S L,Gordon K A,Allemang B,Jewell S,Papsin B C
Abstract
AbstractObjective:To compare stimulation parameters of peri-modiolar and anti-modiolar electrode arrays using two surgical approaches.Methods:Impedance, stimulation thresholds, comfortably loud current levels, electrically evoked compound action potential thresholds and electrically evoked stapedial reflex thresholds were compared between 2 arrays implanted in the same child at 5 time points: surgery, activation/day 1, week 1, and months 1 and 3. The peri-modiolar array was implanted via cochleostomy in all children (n = 64), while the anti-modiolar array was inserted via a cochleostomy in 43 children and via the round window in 21 children.Results:The anti-modiolar array had significantly lower impedance, but required higher current levels to elicit thresholds, comfort, electrically evoked compound action potential thresholds and electrically evoked stapedial reflex thresholds than the peri-modiolar array across all time points, particularly in basal electrodes (p < 0.05). The prevalence of open electrodes was similar in anti-modiolar (n = 5) and peri-modiolar (n = 3) arrays.Conclusion:Significant but clinically acceptable differences in stimulation parameters between peri-modiolar and anti-modiolar arrays persisted four months after surgery in children using bilateral cochlear implants. The surgical approach used to insert the anti-modiolar array had no overall effect on outcomes.
Publisher
Cambridge University Press (CUP)
Subject
Otorhinolaryngology,General Medicine
Cited by
5 articles.