Abstract
AbstractLead cut-outs used to shape fields in kilovoltage radiotherapy can increase the surface dose on the patient. The physical processes leading to increased surface doses are summarised, and an empirical investigation of the efficacy of various coatings in reducing the skin dose generated by secondary electrons released in the lead during irradiation is presented, based on measurements using a thin window parallel plate ionisation chamber in 135 kVp and 225 kVp beams from a Pantak DXT-300 kilovoltage therapy unit. A new flexible coating for lead cut-outs has been formulated and tested. This coating, which is a combination of Copydex and emulsion paint, has been shown to be effective in reducing the skin dose generated by secondary electrons released in the lead during irradiation. The coating is easy to clean, and its inherent elasticity prevents cracking of the coating in clinical use. Its only disadvantage is that rough handling, or contact with sharp objects, can peel the coating at the point of contact.
Publisher
Cambridge University Press (CUP)
Subject
Oncology,Radiology Nuclear Medicine and imaging
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献