An institutional review: dosimetry comparison between simultaneous integrated boost IMRT and VMAT for prostate cancer

Author:

Mukhtar RaheelORCID,Butt Sumera,Rafaye Muhammad Abdur,Iqbal Khalid,Mazhar Sana,Sadaf Tabinda

Abstract

AbstractPurpose:A comparative study was performed about the plan parameters and quality indices between volumetric arc therapy (VMAT) and intensity-modulated radiotherapy (IMRT) for the treatment of high-risk prostate cancer patients. The aim of this retrospective study was to compare the two methods of external beam radiotherapy IMRT and VMAT in terms of plan quality and efficacy.Material and method:Fifteen high-risk prostate patients were planned for radiotherapy using 6 MV photon. Three dose levels were contoured having Planning Tumour Volume 1 (PTV1 = 48 Gy), Planning Tumour Volume 2 (PTV2 = 57.6 Gy) and Planning Tumour Volume 3 (PTV3 = 60 Gy). Setup margins were given using the CHIP trial method. The prescribed PTV3 dose was 60 Gy in 20 fractions which is biologically equivalent to 74 Gy in 37 fractions using α/β = 3. In case of IMRT, seven fixed beam angles 30, 60, 105, 180, 255, 300 and 330 were used and the dose was optimised using the sliding window method. In case of rapid arc technique, one or two full arcs were used for dose optimisation while keeping all the dose constraints and other planning parameters same used in IMRT. The plan evaluation parameters and Organ at risks (OARs) doses were calculated using a dose volume histogram (DVH).Results:The average D2, D5, D95 and PTVmean for PTV3 were 61.22, 61.13, 58.12, 60.00 Gy and 62.41 62.24 59.53 61.12 Gy for IMRT and VMAT, respectively. The averages V60 for bladder and V30 for rectum were 22.81, 25 and 67, 65% for IMRT and VMAT, respectively. The average homogeneity index (HI), conformity index (CI) and gradient index (GI) were 1.04, 1.4833, 14.79 and 1.04, 1.704, 7.89 for IMRT and VMAT, respectively.Conclusion:VMAT takes less dose-delivery time and lesser number of monitoring units than IMRT, thus it compensates the intrafractional movements during dose delivery. The Dose GI in VMAT was much better than IMRT. This indicates sharper dose fall off near the normal tissue. No other major differences were observed in terms of plan evaluation parameters between IMRT and VMAT techniques. So, we conclude that VMAT technique is more efficient than IMRT in terms of plan quality and dose delivery.

Publisher

Cambridge University Press (CUP)

Subject

Oncology,Radiology, Nuclear Medicine and imaging

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3