Abstract
Abstract
Introduction
This study aimed to investigate the uncertainty in organ delineation of low-dose computed tomography (CT) images using a high-strength iterative reconstruction (IR) during radiotherapy planning for the treatment of prostate cancer.
Methods
Two CT datasets were prepared with different dose levels by adjusting the reconstruction slice thickness. Two observers independently delineated the prostate, seminal vesicles, bladder and rectum on both images without referring to other modality images. The delineated organ volumes were compared between both images. Observer delineation variability was assessed using Dice similarity coefficient (DSC) and mean distance to agreement.
Results
No significant differences regarding the delineated organ volumes were observed between the low- and standard-dose images for all organs. Regarding inter-observer variability, the DSC was relatively high for both images, whereas mean distance to agreement was not significantly different between images (p > 0·05 for all). Intra-observer variability for each observer showed high DSC (>0·8 and >0·9 for seminal vesicles and other organs, respectively) but no significant differences in the mean distance to agreement (p > 0·05 for all).
Conclusions
Our results indicate that low-dose CT images with high-strength IR would be available for organ delineation in the radiotherapy treatment planning for prostate cancer.