Radiotherapy couches: is kevlar an obstacle? Attenuation study of three different tabletops

Author:

Silva Célia,Mateus Dalila,Vieira Sandra,Rodrigues Milton,Eiras Margarida,Greco Carlo

Abstract

AbstractIntroductionTreatment tabletops are usually made of carbon fibre due to its high mechanical strength and rigidity, low specific density, extremely light and regularly considered radiotranslucent. Our clinic acquired a Calypso 4D Localization System where electromagnetic (EM) frequencies to detect implanted transponders in the patient are used. Carbon fibre is an electrical conductive material which interferes with EM frequencies. Therefore, in order to be able to use the Calypso System the carbon fibre tabletop in the treatment room must be replaced. It is our goal to determine the attenuation of the new, non-carbon fibre, tabletop in treatment delivery.Materials and MethodsMeasurements were performed using an ionisation chamber inserted in a slab phantom positioned at the isocenter for 6, 10 MV, 6 and 10 flattening filter free (FFF) MV photon beams. These measurements were performed with and without tabletop for 0°, 30° and 60° beam angle for a True Beam STx linac, for 5×5 cm2 and 10×10 cm2 field size beams. The attenuation was calculated for each measurement for each tabletop.ResultsAt 0° incidence on the Exact IGRT Couch, the measured attenuation for 10×10 cm2 was 2·8 and 2·1% for 6 and 10 MV beams, respectively. For the same field size was measured 3·3 and 2·6% attenuation for 6 and 10 FFF MV beams, respectively. At the same incidence and regarding the other tabletops, the calculated attenuation is lower. For 10×10 cm2 field, there is 2·0, 1·4, 2·1 and 2·6% attenuation for 6, 10 MV, 6 and 10 FFF MV energy beams on the kVueTM Universal Couch. For the KvueTM Calypso® Couch 10×10 cm2 irradiation field, the measurements were 1·6, 1·3, 1·9 and 1·5%, respectively. This tendency is observed for all gantry angles.DiscussionThe attenuation outputs were definitely higher for the Varian Exact IGRT Couch when compared with the kVue tabletops. The attenuation measurements for the kVue tabletops were closer to each other. Nevertheless kVueTM Calypso® Varian tabletop showed smaller mean attenuation of the beams than kVueTM Universal Tip Insert for all measurements.ConclusionsThere was no loss in treatment quality administration due to beam attenuation in the tabletop when tabletops were exchanged because of Calypso system integration. There is no need to change between kVue tabletops whenever there is a regular treatment or a Calypso System guided treatment.

Publisher

Cambridge University Press (CUP)

Subject

Oncology,Radiology Nuclear Medicine and imaging

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3