The effect of CT reconstruction filter selection on Hounsfield units in radiotherapy treatment planning

Author:

Nhila OussamaORCID,Talbi Mohammed,El Mansouri M’hamed,Youssoufi Moulay Ali,Erraoudi Morad,Chakir El Mahjoub,Azougagh Mohamed

Abstract

Abstract Introduction: This work aims to evaluate the effect of Hitachi 16-slice scanner reconstruction filters on Hounsfield unit (HU) variations. In the literature, there is a lack of information from a wide variety of scanners in this regard. In addition, not all studies have investigated the effect of reconstruction filters on HU in an exhaustive way. Methods: The computerised imaging reference system electron density phantom (model 062M) was scanned with different substitute materials of different density from Hitachi 16-slice computed tomography. The raw images were obtained with four tube voltage settings: 80 kVp, 100 kVp, 120 kVp and 140 kVp. The raw images for each energy level were then reconstructed using different reconstruction filters. Results: The HU values of dense bone were significantly different when changing the reconstruction filters without beam hardening correction (BHC). Nevertheless, when selecting the BHC, this variation decreases heavily for 80 kVp and decreases slightly for 140 kVp, but it remains outside the tolerance of ±50 HU. However, for 100 kVp and 120 kVp, the differences in HU values become within the tolerances indicated for dense bone. Conclusions: Changing image reconstruction filters during a dosimetric scan had a significant effect on HU in dense bone. Therefore, it is recommended to evaluate this effect during the commissioning phase. As a result, this study provides a methodology to comprehensively investigate the effect of reconstruction filters on HU.

Publisher

Cambridge University Press (CUP)

Subject

Oncology,Radiology, Nuclear Medicine and imaging

Reference30 articles.

1. Impact of Tube Voltage on Radiation Dose (CTDI) and Image Quality at Chest CT Examination

2. Clinical evaluation of a novel CT image reconstruction algorithm for direct dose calculations

3. 29. Kirwin, S , Langmack, K , Nightingale, A. Effect of CT reconstruction kernel and post-processing filter on Hounsfield number constancy in radiotherapy treatment planning. Inst Phys Eng Med SCOPE 2004. https://ctug.org.uk/meet04-10-14/recon_paramters_ct_number.pdf

4. A comparative study based on image quality and clinical task performance for CT reconstruction algorithms in radiotherapy

5. Computed tomography imaging parameters for inhomogeneity correction in radiation treatment planning;Das;J Med Phys,2016