Abstract
AbstractBackground and purposeThis study reported the justification and selection of acceptable γ criteria with respect to low (6 MV) and high (15 MV) photon beams for intensity-modulated radiation therapy quality assurance (IMRT QA) using the Gafchromic external beam therapy 3 (EBT3) film.Materials and methodsFive-field step-and-shoot IMRT was used to treat 16 brain IMRT patients using the dual-energy DHX-S linear accelerator (Varian Medical System, Palo Alto, CA, USA). Dose comparisons between computed values of the treatment planning system (TPS) and Gafchromic EBT3 film were evaluated based on γ analysis using the Film QA Pro software. The dose distribution was analysed with gamma area histograms (GAHs) generated using different γ criteria (3%/2 mm, 3%/3 mm and 5%/3 mm) for the 6 and 15 MV photon beams, to optimise the best distance-to-agreement (DTA) criteria with respect to the beam energy.ResultsFrom the comparison between the dose distributions acquired from the TPS and EBT3 film, a DTA criterion of 3%/2 mm showed less dose differences (DDs) with passing rates up to 93% for the 6 MV photon beams, while for the 15 MV a relaxed DTA criterion of 5%/3 mm was consistent with the DD acceptability criteria with a 95% passing rate.ConclusionsOur results suggested that high-energy photon beams required relaxed DTA criteria for the brain IMRT QA, while low-energy photon beams showed better results even with tight DTA criteria.
Publisher
Cambridge University Press (CUP)
Subject
Oncology,Radiology Nuclear Medicine and imaging
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献