Abstract
AbstractAim:To modify the final dose delivered to superficial tissues and to modulate dose distribution near irradiated surface, different boluses are used. Air gaps often form under the bolus affecting dose distribution. This study aimed to evaluate the effect of an air gap under the bolus radiation on dose delivery.Materials and methods:To evaluate the impact of the air gap, both helical tomotherapy (HT) and direct tomotherapy (DT) were performed in a simulation study.Results:The maximum dose to bolus in DT plans was bigger than that used in HT plans. The maximum dose delivered to the bolus depended on the air gap size. However, the maximum dose to bolus in all HT plans was within the acceptable value range. Acceptable value was set to up to 107% of the prescription dose. In the simulation performed in this study, the acceptable air gap under bolus was up to 15 mm and below 5 mm in HT and DT plans, respectively.Conclusions:HT technique is a good choice, but DT technique can be also used if the bolus position can be reproduced accurately. Thus, the reproducibility of the bolus position between planning and treatment is very important.