Author:
Haider M.,Hartel P.,Müller H.,Uhlemann S.,Zach J.
Abstract
AbstractFor the transmission electron aberration-corrected microscope (TEAM) initiative of five U.S. Department of Energy laboratories in the United States, a correction system for the simultaneous compensation of the primary axial aberrations, the spherical aberration Cs, and the chromatic aberration Cc has been developed and successfully installed. The performance of the resulting Cc /Cs-corrected TEAM instrument has been investigated thoroughly. A significant improvement of the linear contrast transfer can be demonstrated. The information about the instrument one obtains using Young's fringe method is compared for uncorrected, Cs-corrected, and Cc /Cs-corrected instruments. The experimental results agree well with simulations. The conclusions might be useful to others in understanding the process of image formation in a Cc /Cs-corrected transmission electron microscope.
Publisher
Cambridge University Press (CUP)
Reference31 articles.
1. Elektronenmikroskopische Bilder mit einem nach O. Scherzer sphärisch korrigierten Objektiv;Möllenstedt;Optik,1956
2. Sphärische und chromatische Korrektur von Elektronenlinsen;Scherzer;Optik,1947
3. Phase tomography in the corrected electron microscope
4. Relativistic Hartree–Fock X-ray and electron scattering factors
5. Aberration correction in a low voltage SEM by a multipole corrector
Cited by
98 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献