L-band radiometer design for long-term stability measurement of one-port devices
-
Published:2011-12-14
Issue:1
Volume:4
Page:119-126
-
ISSN:1759-0787
-
Container-title:International Journal of Microwave and Wireless Technologies
-
language:en
-
Short-container-title:Int. j. microw. wirel. technol.
Author:
de la Jarrige Emilie Leynia,Escotte Laurent,Gonneau Eric,Goutoule Jean-Marc
Abstract
The design of a high-stability radiometer is presented in this paper. It is used to characterize the stability of a microwave active cold load at L-band. The two-load radiometer takes advantage of a noise injection technique to improve its sensitivity. A temperature-stabilized enclosure is used to minimize gain and noise temperature fluctuations of the receiver. The radiometer sensitivity is 0.031 K and the gain fluctuations are less than 0.005 dB during 1 day. The total gain and the receiver noise temperature of the radiometer exhibit very small variations (<0.03 dB and 1 K, respectively) on the long term.
Publisher
Cambridge University Press (CUP)
Subject
Electrical and Electronic Engineering
Reference31 articles.
1. Sensibilité des radiotélescopes et récepteurs à corrélation;Blum;Ann. Astrophys.,1959
2. An Active "Cold" Noise Source
3. The Measurement of Thermal Radiation at Microwave Frequencies
4. A high stability Ka-band radiometer for tropospheric water vapor measurements;Tanner;IEEE Aerospace Conf.,2001