Author:
Smith Adam James,Jackimczyk Ken,Horwood Bruce,Christenson Daniel
Abstract
AbstractBackgroundIn the prehospital setting, many providers advocate for video laryngoscopy as the initial method of intubation to improve the likelihood of a successful first attempt. However, bright ambient light can worsen visualization of the video laryngoscope liquid crystal display (LCD).Case ReportA patient involved in a motor vehicle accident was evaluated by an Emergency Medical Services (EMS) crew. Initial endotracheal intubation attempt using video laryngoscopy was aborted after the patient desaturated. The primary reason for the failure was poor visualization of the video laryngoscope LCD, despite attempts to block direct sunlight. Debriefing revealed that the intubating provider was wearing polarized sunglasses.DiscussionBecause LCDs emit polarized light, use of polarized sunglasses may cause the display to appear dark. Thus, the purpose of this Case Report is to raise awareness of a potential safety issue that is likely under-recognized by prehospital providers but can be easily avoided.SmithAJ, JackimczykK, HorwoodB, ChristensonD. Use of polarized sunglasses during video laryngoscopy: a cause of difficult prehospital intubation.Prehosp Disaster Med. 2019;34(1):104–107.
Publisher
Cambridge University Press (CUP)
Subject
Emergency Nursing,Emergency Medicine
Cited by