DC Defibrillators: The Difference Between Selected and Delivered Energy

Author:

Shiner Stephen L.,Gold Martin I.

Abstract

AbstractThere is controversy over the “ideal” electrical energy needed for defibrillation. Furthermore, too massive an electrical shock decreases the possibility of survival by direct damage to an already ailing myocardium and too little energy results in further deterioration of myocardial physiology and metabolism. Therefore, delivery of an erroneous amount of electrical energy decreases the likelihood of successful defibrillation and survival. All 190 defibrillators within 3 medical school hospitals were investigated. Each defibrillator was analyzed at four selected settings ranging from 100-400 Joules (J). Only 29 of 190 defibrillators delivered 100% of the energy selected, while 161 of the 190 (85%) delivered an average of 74% of the energy selected. This discrepancy between selected and delivered energy should be corrected by regulations and standards for manufacturers.

Publisher

Cambridge University Press (CUP)

Reference42 articles.

1. Digital cardiac defibrillator tester

2. Transthoracic ventricular defibrillation in adults.

3. Energy dose for defibrillation;Babbs;N Engl J Med,1978

4. Transthoracic ventricular defibrillation in the adult;Adgey;Med Instrum,1978