Cadnum Jennifer L.,Tomas Myreen E.,Sankar Thriveen,Jencson Annette,Mathew J. Itty,Kundrapu Sirisha,Donskey Curtis J.
Abstract
OBJECTIVETo determine the effect of variation in test methods on performance of an ultraviolet-C (UV-C) room decontamination device.DESIGNLaboratory evaluation.METHODSWe compared the efficacy of 2 UV-C room decontamination devices with low pressure mercury gas bulbs. For 1 of the devices, we evaluated the effect of variation in spreading of the inoculum, carrier orientation relative to the device, type of organic load, type of carrier, height of carrier, and uninterrupted versus interrupted exposures on measured UV-C killing of methicillin-resistant Staphylococcus aureus and Clostridium difficile spores.RESULTSThe 2 UV-C room decontamination devices achieved similar log10 colony-forming unit reductions in the pathogens with exposure times ranging from 5 to 40 minutes. On steel carriers, spreading of the inoculum over a larger surface area significantly enhanced killing of both pathogens, such that a 10-minute exposure on a 22-mm2 disk resulted in greater than 2 log reduction in C. difficile spores. Orientation of carriers in parallel rather than perpendicular with the UV-C lamps significantly enhanced killing of both pathogens. Different types of organic load also significantly affected measured organism reductions, whereas type of carrier, variation in carrier height, and interrupted exposure cycles did not.CONCLUSIONSVariation in test methods can significantly impact measured reductions in pathogens by UV-C devices during experimental testing. Our findings highlight the need for standardized laboratory methods for testing the efficacy of UV-C devices and for evaluations of the efficacy of short UV-C exposure times in real-world settings.Infect Control Hosp Epidemiol 2016;37:555–560
Publisher
Cambridge University Press (CUP)
Subject
Infectious Diseases,Microbiology (medical),Epidemiology
Cited by
36 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献