Abstract
A commercial Empyrean X-ray diffractometer was adapted for combined grazing incidence X-ray fluorescence analysis (GIXRF) measurements with X-ray reflectivity (XRR) measurements. An energy-dispersive silicon drift detector was mounted and integrated in the angle-dependent data acquisition of the Empyrean. Different monochromator/X-ray optics units have been compared with the values obtained by the Atominstitut GIXRF + XRR spectrometer. Data evaluation was performed by JGIXA, a special software for combined GIXRF + XRR data fitting, developed at Atominstitut. A sample consisting of a ~50 nm nickel layer on a silicon substrate was used to compare the performance criteria (i.e. divergence and intensity) of the incident beam optics. An Empyrean X-ray diffractometer was successfully refitted to measure both GIXRF and XRR data.
Publisher
Cambridge University Press (CUP)
Subject
Condensed Matter Physics,Instrumentation,General Materials Science,Radiation
Cited by
1 articles.
订阅此论文施引文献