Walther Thomas,Stegmann Heiko
Abstract
Experimental results from the first monochromated and
aberration-corrected scanning transmission electron microscope operated at
200 kV are described. The formation of an electron probe with a diameter
of less than 0.2 nm at an energy width significantly under 0.3 eV and its
planned application to the chemical analysis of nanometer-scale structures
in materials science are described. Both energy and spatial resolution
will benefit from this: The monochromator improves the energy resolution
for studies of energy loss near edge structures. The Cs corrector allows
formation of either a smaller probe for a given beam current or yields, at
fixed probe size, an enhanced beam current density using a larger
condenser aperture. We also point out another advantage of the combination
of both components: Increasing the convergence angle by using larger
condenser apertures in an aberration-corrected instrument will enlarge the
undesirable chromatic focus spread. This in turn influences spatial
resolution. The effect of polychromatic probe tails is proportional to the
product of convergence angle, chromatic aberration constant, and energy
spread. It can thus be compensated for in our new instrument by decreasing
the energy width by the same factor as the beam convergence is increased
to form a more intense probe. An alternative in future developments might
be hardware correction of the chromatic aberration, which could eliminate
the chromatic probe spread completely.
Publisher
Cambridge University Press (CUP)
Reference25 articles.
1. Hutchison, J.L. , Titchmarsh, J.M. , Cockayne, D.J.H. , Doole, R.C. , Hetherington, C.J.D. , Kirkland, A.I. , & Sawada, H. (2005).A versatile double aberration-corrected, energy-filteredHREM/STEM for materials science.Ultramicroscopy 103,7–15.
2. Rose, H. (1990b).Outline of a spherically corrected semiaplanatic medium-voltagetransmission electron-microscope.Optik 85,19–24.
3. Walther, T. (2006).A new experimental procedure to quantify annular dark field imagesin scanning transmission electron microscopy.J Microsc 221,137–144.
4. Mook, H.W. & Kruit, P. (1999).Optics and design of the fringe field monochromator for a Schottkyfield emission gun.Nucl Instrum Methods Phys Res A 427,109–120.
5. Kreibig, U. & Zacharias, P. (1970).Surface plasma resonances in small spherical silver and goldparticles.Z Phys 213,128–143.
Cited by
20 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献