Sadler R. Mark,Goodwin John
Abstract
ABSTRACT:The contribution of various electroencephalographic electrodes in detecting spikes from patients with seizures of suspected anterior temporal origin was prospectively studied with a standard protocol. The following electrodes were studied: International Standard 10-20 positions F7-8 and A1-2, sphenoidal (SP), nasopharyngeal (NP), anterior temporal (Tl-2), mandibular notch surface (MNS), and mandibular notch subdermal (MNSD). Twenty patients were recorded of whom 16 demonstrated anterior temporal spikes. There was no difference in the number of spikes detected by SP, MNS, MNSD, or Tl-2 electrodes (p < 0.05); however these electrodes detected significantly more spikes than NP, F7-8, or A1-2. The SP electrode recorded spikes of highest amplitude (p < 0.05). We conclude that for patients suspected of having seizures of anterior temporal origin, (1) a substantial number of spikes will be missed if only the International Standard electrode system is employed; (2) in comparison to SP electrodes the noninvasive and easily applied MNS or Tl-2 electrodes will detect almost all spikes and should be used in outpatient EEG recordings; (3) NP electrodes provide no information that cannot be obtained by more reliable and better tolerated electrodes.
Publisher
Cambridge University Press (CUP)
Subject
Clinical Neurology,Neurology,General Medicine
Cited by
65 articles.