Abstract
Cochlear implantation is currently the most effective treatment modality for severe to profound sensorineural hearing loss. Over the past few years, at the Department of Otolaryngology, Cheng Hsin General Hospital (Taipei, Taiwan), cochlear implant devices have been switched on within 24 hours of their implantation. Differences in impedance evolution after early switch-on for different devices have not been previously discussed. The present study aimed to investigate the impedance evolution of one device and the factors influencing this after early activation. Results are compared to published results of other devices. A total of 16 patients who received Advanced BionicsTM devices and had early activation within 24 hours of implantation, were included in the study. Impedance telemetry was recorded intraoperatively and postoperatively at 1 day, 1 week, 2 weeks, 4 weeks and 8 weeks. A stepwise increase was observed in the impedance evolution. To the best of our knowledge, the present study is the first to investigate the impedance evolution of the different devices after early switch-on within 24 hours of implantation and its influencing factors. Further research with a longitudinal design to compare the differences in electrode impedances between patients activated early versus those activated after a few weeks will be necessary for the disclosure of the underlying mechanisms.
Funder
Cheng Hsin General Hospital Foundation
Publisher
Public Library of Science (PLoS)
Cited by
12 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献