Validation of a new wearable device for type 3 sleep test without flowmeter

Author:

Contini Mauro,Sarmento AntonioORCID,Gugliandolo Paola,Leonardi Alessandra,Longinotti-Buitoni Gianluigi,Minella Camilla,Vignati Carlo,Mapelli MassimoORCID,Aliverti AndreaORCID,Agostoni Piergiuseppe

Abstract

Background Ventilation monitoring during sleep is performed by sleep test instrumentation that is uncomfortable for the patients due to the presence of the flowmeter. The objective of this study was to evaluate if an innovative type 3 wearable system, the X10X and X10Y, is able to correctly detect events of apnea and hypopnea and to classify the severity of sleep apnea without the use of a flowmeter. Methods 40 patients with sleep disordered breathing were analyzed by continuous and simultaneous recording of X10X and X10Y and another certified type 3 system, SOMNOtouch, used for comparison. Evaluation was performed in terms of quality of respiratory signals (scores from 1, lowest, to 5, highest), duration and classification of apneas, as well as identification and duration of hypopneas. Results 580 periods were evaluated. Mean quality assigned score was 3.37±1.42 and 3.25±1.35 for X10X and X10Y and SOMNOtouch, respectively. The agreement between the two systems was evaluated with grades 4 and 5 in 383 out of 580 cases. A high correlation (r2 = 0.921; p<0.001) was found between the AHI indexes obtained from the two systems. X10X and X10Y devices were able to correctly classify 72.3% of the obstructive apneas, 81% of the central apneas, 61.3% of the hypopneas, and 64.6% of the mixed apneas when compared to SOMNOtouch device. Conclusion The X10X and X10Y devices are able to provide a correct grading of sleep respiratory disorders without the need of a nasal cannula for respiratory flow measurement and can be considered as a type 3 sleep test device for screening tests.

Funder

L.I.F.E. Italia S.r.l. and L.I.F.E. Corporation S.A.

Publisher

Public Library of Science (PLoS)

Subject

Multidisciplinary

Cited by 8 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Wearable Technologies for Healthy Ageing: Prospects, Challenges, and Ethical Considerations;Journal of Frailty & Aging;2024

2. Design and Evaluation of a Wearable Single-Lead ECG for Continuous Monitoring;2023 IEEE International Conference on Metrology for eXtended Reality, Artificial Intelligence and Neural Engineering (MetroXRAINE);2023-10-25

3. Dysregulation of ventilation at day and night time in heart failure;European Journal of Preventive Cardiology;2023-10

4. Influence-Based Nano Fuzzy Swarm Oxygen Deficiency Detection and Therapy;IEEE Transactions on Systems, Man, and Cybernetics: Systems;2023-08

5. Influence of Photoplethysmogram Signal Quality on Pulse Arrival Time during Polysomnography;Sensors;2023-02-16