Abstract
Fine bubbles (FBs) have attracted significant attention in several research fields. Although some reports have argued that FB dispersion is useful as an oxygen (gas) carrier, only a few reports have examined its properties as an oxygen carrier using experimental data. As one of the reasons for this, there are no standard methods for measuring the oxygen content in FB dispersions. Conventional oxygen measurement methods have certain drawbacks in accuracy or speed; thus, it is difficult to use oxygen content as the primary outcome. In this study, we introduce a Clark-type polarographic oxygen electrode device (OXYG1-PLUS) for oxygen measurement, allowing the dilution of FB dispersion without the influence of ambient air and the adhesion of FBs on the electrode surface due to its special shape. First, the accuracy of our dilution method was evaluated using pure water as a sample, and it was confirmed that our method could measure with an accuracy of ±0.5 mg/L from the results with conventional dissolved oxygen meters. Second, the oxygen content in FB dispersion was evaluated with our method and a chemical titration method (Winkler’s method), and it was found that our method could measure the oxygen content in FB dispersions quantitively. This method satisfies the easiness (4 steps) and quickness (within 8 min) for a wide range of oxygen contents (0 to 332 mg/L, theoretical range) with low coefficient variation (< 4.7%) and requires a small sample volume (50–500 μL); thus, it is a useful method for measuring the oxygen in FB dispersions.