Effect of optical blurring of X-ray source on breast tomosynthesis image quality: Modulation transfer function, anatomical noise power spectrum, and signal detectability perspectives

Author:

Lee ChangwooORCID,Baek JongdukORCID

Abstract

We investigated the effect of the optical blurring of X-ray source on digital breast tomosynthesis (DBT) image quality using well-designed DBT simulator and table-top experimental systems. To measure the in-plane modulation transfer function (MTF), we used simulated sphere phantom and Teflon sphere phantom and generated their projection data using two acquisition modes (i.e., step-and-shoot mode and continuous mode). After reconstruction, we measured the in-plane MTF using reconstructed sphere phantom images. In addition, we measured the anatomical noise power spectrum (aNPS) and signal detectability. We constructed simulated breast phantoms with a 50% volume glandular fraction (VGF) of breast anatomy using the power law spectrum and inserted spherical objects with 1 mm, 2 mm, and 5 mm diameters as breast masses. Projection data were acquired using two acquisition modes, and in-plane breast images were reconstructed using the Feldkamp-Davis-Kress (FDK) algorithm. For the experimental study, we used BR3D breast phantom with 50% VGF and obtained projection data using a table-top experimental system. To compare the detection performance of the two acquisition modes, we calculated the signal detectability using the channelized Hotelling observer (CHO) with Laguerre-Gauss (LG) channels. Our results show that spatial resolution of in-plane image in continuous mode was degraded due to the optical blurring of X-ray source. This blurring effect was reflected in aNPS, resulting in large β values. From a signal detectability perspective, the signal detectability in step-and-shoot mode is higher than that in continuous mode for small spherical signals but not large spherical signals. Although the step-and-shoot mode has disadvantage in terms of scan time compared to the continuous mode, scanning in step-and-shoot mode is better for detecting small signals, indicating that there is a tradeoff between scan time and image quality.

Funder

R&D Programs

National Research Foundation

National Research Foundtion

Korea Medical Device Development Fund

Publisher

Public Library of Science (PLoS)

Subject

Multidisciplinary

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3