Correlation of two different devices for the evaluation of primary implant stability depending on dental implant length and bone density: An in vitro study

Author:

Lee JungwonORCID,Lim Young-Jun,Ahn Jin-Soo,Kim BongjuORCID,Baek Yeon-WhaORCID,Lim Bum-Soon

Abstract

Non-invasive objective implant stability measurements are needed to determine the appropriate timing of prosthetic fitting after implant placement. We compared the early implant stability results obtained using resonance frequency analysis (RFA) and damping capacity analysis (DCA) depending on the implant length and bone density. Total 60, 4.0 mm diameter implants of various lengths (7.3 mm, 10 mm, and 13 mm) were used. In Group I, low-density bone was described using 15 PCF (0.24 g/cm3) polyurethane bone blocks, and in Group II, 30 PCF (0.48 g/cm3) polyurethane bone blocks were used to describe medium density bone. RFA was performed using an Osstell® Beacon+; DCA was performed using Anycheck®. Measurements were repeated five times for each implant. Statistical significance was set at P <0.05. In Group I, bone density and primary implant stability were positively correlated, while implant length and primary implant stability were positively correlated. In Group II, the implant stability quotient (ISQ) and implant stability test (IST) values in did not change significantly above a certain length. Primary implant stability was positively correlated with bone density and improved with increasing implant length at low bone densities. Compared with the Osstell® Beacon+, the simplicity of Anycheck® was easy to use and accessible.

Funder

Korea Medical Device Development Fund

Publisher

Public Library of Science (PLoS)

Reference30 articles.

1. Accuracy of biplanar linear radiography versus conventional radiographs when used for lower limb and implant measurements;CXY Chua;Archives of Orthopaedic and Trauma Surgery,2021

2. Marginal bone level changes and implant stability after loading are not influenced by baseline microstructural bone characteristics: 1‐year follow‐up;DR Dias;Clinical oral implants research,2016

3. Quantitative determination of the stability of the implant‐tissue interface using resonance frequency analysis;N Meredith;Clinical oral implants research,1996

4. Determination of the optimum torque to tighten the smartpegs of magnetic resonance frequency analyses devices: an ex vivo study;O Geckili;Journal of Oral Implantology,2015

5. Does the manual insertion torque of smartpegs affect the outcome of implant stability quotients (ISQ) during resonance frequency analysis (RFA)?;I Kästel;International Journal of Implant Dentistry,2019

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3