Field assessment of the operating procedures of a semi-quantitative G6PD Biosensor to improve repeatability of routine testing

Author:

Sadhewa ArkashaORCID,Chaudhary Alina,Panggalo Lydia V.,Rumaseb Angela,Adhikari Nabaraj,Adhikari Sanjib,Rijal Komal RajORCID,Banjara Megha Raj,Price Ric N.ORCID,Thriemer Kamala,Ghimire Prakash,Ley BenediktORCID,Satyagraha Ari Winasti

Abstract

In remote communities, diagnosis of G6PD deficiency is challenging. We assessed the impact of modified test procedures and delayed testing for the point-of-care diagnostic STANDARD G6PD (SDBiosensor, RoK), and evaluated recommended cut-offs. We tested capillary blood from fingerpricks (Standard Method) and a microtainer (BD, USA; Method 1), venous blood from a vacutainer (BD, USA; Method 2), varied sample application methods (Methods 3), and used micropipettes rather than the test’s single-use pipette (Method 4). Repeatability was assessed by comparing median differences between paired measurements. All methods were tested 20 times under laboratory conditions on three volunteers. The Standard Method and the method with best repeatability were tested in Indonesia and Nepal. In Indonesia 60 participants were tested in duplicate by both methods, in Nepal 120 participants were tested in duplicate by either method. The adjusted male median (AMM) of the Biosensor Standard Method readings was defined as 100% activity. In Indonesia, the difference between paired readings of the Standard and modified methods was compared to assess the impact of delayed testing. In the pilot study repeatability didn’t differ significantly (p = 0.381); Method 3 showed lowest variability. One Nepalese participant had <30% activity, one Indonesian and 10 Nepalese participants had intermediate activity (≥30% to <70% activity). Repeatability didn’t differ significantly in Indonesia (Standard: 0.2U/gHb [IQR: 0.1–0.4]; Method 3: 0.3U/gHb [IQR: 0.1–0.5]; p = 0.425) or Nepal (Standard: 0.4U/gHb [IQR: 0.2–0.6]; Method 3: 0.3U/gHb [IQR: 0.1–0.6]; p = 0.330). Median G6PD measurements by Method 3 were 0.4U/gHb (IQR: -0.2 to 0.7, p = 0.005) higher after a 5-hour delay compared to the Standard Method. The definition of 100% activity by the Standard Method matched the manufacturer-recommended cut-off for 70% activity. We couldn’t improve repeatability. Delays of up to 5 hours didn’t result in a clinically relevant difference in measured G6PD activity. The manufacturer’s recommended cut-off for intermediate deficiency is conservative.

Funder

Department of Foreign Affairs and Trade, Australian Government

Charles Darwin University

Division of Teaching of the Menzies School of Health Research

Publisher

Public Library of Science (PLoS)

Reference36 articles.

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3