Abstract
Purpose
To measure needle insertion force and change in intraocular pressure (IOP) in real-time during intravitreal injection (IVI). The effects of needle size, insertion speed, and injection rate to IOP change were investigated.
Methods
Needle insertion and fluid injection were performed on 90 porcine eyeballs using an automatic IVI device. The IVI conditions were divided according to needle sizes of 27-gauge (G), 30G, and 33G; insertion speeds of 1, 2, and 5 mm/s; and injection rates of 0.01, 0.02, and 0.05 mL/s. Insertion force and IOP were measured in real-time using a force sensor and a pressure transducer.
Results
The peak IOP was observed when the needle penetrated the sclera; the average IOP elevation was 96.3, 67.1, and 59.4 mmHg for 27G, 30G, and 33G needles, respectively. An increase in insertion speed caused IOP elevation at the moment of penetration, but this effect was reduced as needle size decreased: 109.8–85.9 mmHg in 27G for 5–1 mm/s (p = 0.0149) and 61.8–60.7 mmHg in 33G for 5–1 mm/s (p = 0.8979). Injection speed was also related to IOP elevation during the stage of drug injection: 16.65 and 11.78 mmHg for injection rates of 0.05 and 0.01 mL/s (p < 0.001).
Conclusion
The presented data offers an understanding of IOP changes during each step of IVI. Slow needle insertion can reduce IOP elevation when using a 27G needle. Further, the injection rate must be kept low to avoid IOP elevations during the injection stage.