Multi-site, multi-platform comparison of MRI T1 measurement using the system phantom

Author:

Keenan Kathryn E.ORCID,Gimbutas Zydrunas,Dienstfrey Andrew,Stupic Karl F.,Boss Michael A.ORCID,Russek Stephen E.,Chenevert Thomas L.,Prasad P. V.,Guo Junyu,Reddick Wilburn E.ORCID,Cecil Kim M.ORCID,Shukla-Dave Amita,Aramburu Nunez David,Shridhar Konar Amaresh,Liu Michael Z.,Jambawalikar Sachin R.,Schwartz Lawrence H.,Zheng Jie,Hu Peng,Jackson Edward F.

Abstract

Recent innovations in quantitative magnetic resonance imaging (MRI) measurement methods have led to improvements in accuracy, repeatability, and acquisition speed, and have prompted renewed interest to reevaluate the medical value of quantitative T1. The purpose of this study was to determine the bias and reproducibility of T1 measurements in a variety of MRI systems with an eye toward assessing the feasibility of applying diagnostic threshold T1 measurement across multiple clinical sites. We used the International Society of Magnetic Resonance in Medicine/National Institute of Standards and Technology (ISMRM/NIST) system phantom to assess variations of T1 measurements, using a slow, reference standard inversion recovery sequence and a rapid, commonly-available variable flip angle sequence, across MRI systems at 1.5 tesla (T) (two vendors, with number of MRI systems n = 9) and 3 T (three vendors, n = 18). We compared the T1 measurements from inversion recovery and variable flip angle scans to ISMRM/NIST phantom reference values using Analysis of Variance (ANOVA) to test for statistical differences between T1 measurements grouped according to MRI scanner manufacturers and/or static field strengths. The inversion recovery method had minor over- and under-estimations compared to the NMR-measured T1 values at both 1.5 T and 3 T. Variable flip angle measurements had substantially greater deviations from the NMR-measured T1 values than the inversion recovery measurements. At 3 T, the measured variable flip angle T1 for one vendor is significantly different than the other two vendors for most of the samples throughout the clinically relevant range of T1. There was no consistent pattern of discrepancy between vendors. We suggest establishing rigorous quality control procedures for validating quantitative MRI methods to promote confidence and stability in associated measurement techniques and to enable translation of diagnostic threshold from the research center to the entire clinical community.

Funder

National Institutes of Health

Publisher

Public Library of Science (PLoS)

Subject

Multidisciplinary

Reference39 articles.

1. Clinical NMR imaging of the brain: 140 cases;GM Bydder;AJR American journal of roentgenology,1982

2. MR imaging: possibility of tissue characterization of brain tumors using T1 and T2 values;M Komiyama;AJNR American journal of neuroradiology,1987

3. Tissue characterization with T1, T2, and proton density values: results in 160 patients with brain tumors;M Just;Radiology,1988

4. Tissue characterization of intracranial tumors by MR imaging;L Kjaer;In vivo evaluation of T1- and T2-relaxation behavior at 1.5 T. Acta radiologica,1991

5. T1, T2 and proton density measurements in the grading of cerebral gliomas;S Newman;European Radiology,1993

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3