Abstract
Current methods for detection of mycotoxin in feed are time-consuming and tedious. An up-converting phosphor technology-based lateral flow (UPT-LF) assay system is a new emerging technique for analytes detection. The aim of this study was to compare the performance of UPT-LF, an enzyme-linked immunosorbent assay (ELISA) and high-performance liquid chromatography-tandem mass spectrometry (HPLC-MS/MS) for detecting aflatoxin B1 (AFB1), zearalenone (ZEN) and deoxynivalenol (DON) in feed. The results showed that the use of UPT-LF for AFB1, ZEN and DON detection exhibited the following: limits of detection of 3, 50 and 200 μg/kg; average recoveries of 104.39%, 102.94% and 103.65%; and precision of 13.96%, 13.71% and 12.56%; respectively. UPT-LF required 45 min to determine one mycotoxin and 1.5 h to determine three mycotoxins in a sample, which took the shortest time. Besides, there were positive correlations between the UPT-LF, ELISA and HPLC/MS/MS methods. In conclusion, UPT-LF can be used to detect and quantify AFB1, ZEN and DON in feed samples. Though the sensitivity, accuracy and precision of UPT-LF are inferior to those of HPLC-MS/MS and ELISA, the UPT-LF assay is the most convenient and rapid technique for on-site detection among the three methods.
Funder
National Key Research and Development Program of China
Major Scientific and Technological Innovation Project of Shandong Province
Modern Agricultural Industry Technology of Shandong Province
Publisher
Public Library of Science (PLoS)
Cited by
8 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献