Abstract
Objective
To compare the angle of deviation measured from Photo-Hirschberg testing and Krimsky testing, with that from an alternate prism cover test (APCT) in strabismus patients.
Methods
A cross-sectional study was conducted in Songklanagarind Hospital, Thailand. Thirty-three strabismus patients were photographed for analysis by Photo-Hirschberg testing using computer software. The corneal light reflex displacement, converted into prism diopter (PD), was compared to the angle of deviation measured with APCT. Twenty-eight strabismus patients were tested with the Krimsky test. Data were analyzed using Pearson correlation and paired t-tests. The study excluded 4 intermittent exotropia cases, 1 intermittent esotropia case and 2 which cases missing data for krimsky test.
Results
The mean±SD of the deviation angle, measured by APCT with a fixation target at 30 cm and 6 m; were 48.09±16.34PD and 47.82±15.73 PD, respectively. At 1 m, the difference in the angle of deviation measured from APCT and the Photo-Hirschberg test within 10 PD were 58.8% and 63.6%, for ET and XT, respectively. The difference in the angle of deviation measured from APCT and Krimsky tests within 10 PD in ET and XT were 86.7% and 80.0%, respectively. At 4 m, the difference in angle of deviation measured from APCT and Photo-Hirschberg tests within 10 PD in ET and XT were 58.8% and 54.5%, respectively; whereas, the difference in the angle of deviation measured from APCT and Krimsky tests within 10 PD in ET and XT were 80.0% and 70.0%, respectively.
Conclusion
The reliability of Krimsky test was better than Photo-Hirschberg test for measuring an angle of deviation.
Funder
Faculty of Medicine, Prince of Songkla University
Publisher
Public Library of Science (PLoS)
Reference22 articles.
1. Tests for detecting strabismus in children aged 1 to 6 years in the community;S Hull;Cochrane Database Syst Rev,2017
2. The calibration of the corneal light reflex to estimate the degree of an angle of deviation;S Tengtrisorn;J Med Assoc Thai,2015
3. Biostatistics 104: correlational analysis;YH Chan;Singap Med J,2003
4. Field testing of the plusoptiX S04 photoscreener;BW Arthur;Journal of AAPOS,2009
5. Calibration of Hirschberg test photographs under clinical conditions;PA DeRespinis;Ophthalmology,1989
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献