Abstract
The quadrilateral anterior cervical plate (ACP) is used extensively in anterior cervical discectomy and fusion (ACDF) to reconstruct the stability of the cervical spine and prevent cage subsidence. However, there have been no comparison studies on the biomechanical performance of quadrilateral ACP and triangular ACP. The objective of this study is to investigate the functional outcomes of quadrilateral ACP and triangular ACP usage in ACDF surgery. In this study, a finite element model of intact C1-C7 segments was established and verified. Additionally, two implant systems were built; one using triangle anterior cervical plates (TACP) and another using quadrilateral orion anterior cervical plate (QACP). Both models were then compared in terms of their postoperative biomechanical performance, under normal and excessive motion. Compared to QACP, the peak stress of the TACP screws and plates occurred at 359.2 MPa and 97.2 MPa respectively and were the highest during over extension exercises. Alternately, compared to TACP, the endplate peak stress and the cage displacement of QACP were the largest at over extension, with values of 7.5 MPa and 1.2 mm, respectively. Finally, the average stress ratio of bone grafts in TACP was relatively high at 31.6%. In terms of biomechanical performance, TACP can share the load more flexibly and reduce the risks of cage subsidence and slippage but the screws have high peak stress value, thereby increasing the risk of screw slippage and fracture. This disadvantage must be considered when designing a TACP based implant for a potential patient.
Funder
National Natural Science Foundation of China
Publisher
Public Library of Science (PLoS)
Cited by
6 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献