Abstract
Changes in limb volume and shape among transtibial amputees affects socket fit and comfort. The ability to accurately measure residual limb volume and shape and relate it to comfort could contribute to advances in socket design and overall care. This work designed and validated a novel 3D laser scanner that measures the volume and shape of residual limbs. The system was designed to provide accurate and repeatable scans, minimize scan duration, and account for limb motion during scans. The scanner was first validated using a cylindrical body with a known shape. Mean volumetric errors of 0.17% were found under static conditions, corresponding to a radial spatial resolution of 0.1 mm. Limb scans were also performed on a transtibial amputee and yielded a standard deviation of 8.1 ml (0.7%) across five scans, and a 46 ml (4%) change in limb volume when the socket was doffed after 15 minutes of standing.
Funder
National Science Foundation
Publisher
Public Library of Science (PLoS)
Reference35 articles.
1. Estimating the prevalence of limb loss in the United States: 2005 to 2050;K. Ziegler-Graham;Archives of physical medicine and rehabilitation,2008
2. Reduction of residual limb volume in people with transtibial amputation;A. T. Tantua;J Rehabil Res Dev,2014
3. Shape and volume change in the transtibial residuum over the short term: preliminary investigation of six subjects;S. G. Zachariah;Journal of Rehabilitation Research & Development,2004
4. Volume changes occurring in postoperative below-knee residual limbs;F. L. Golbranson;J Rehabil Res Dev,1988
5. Residual Limb Volume Change: Systematic Review of Measurement and Management;J. E. Sanders;Journal of Rehabilitation Research & Development,2011