Abstract
Purpose
To develop artificial eye models using 3D printing and to evaluate the correlation between different corneal thicknesses and intraocular pressures (IOPs).
Methods
We designed 7 artificial eye models using a computer-aided design system and fabricated them using 3D printing. Corneal curvature and axial length were based on the Gullstrand eye model. Hydrogels were injected into the vitreous cavity, and seven different corneal thicknesses (200 to 800 μm) were prepared. In this proposed design, we also produced different corneal stiffnesses. A Tono-Pen AVIA tonometer was used by the same examiner to perform five consecutive IOP measurements in each eye model.
Results
Different eye models were ideally created using 3D printing. IOP measurements were successfully performed in each eye model. The corneal thickness was significantly correlated with IOP (R2 = 0.927; 𝑃<0.001).
Conclusion
The 3D-printed eye model is useful for evaluating IOP measurements. This technique might be a promising alternative to the conventional porcine eye model.
Publisher
Public Library of Science (PLoS)
Reference9 articles.
1. 3D-printed hierarchical pillar array electrodes for high-performance semi-artificial photosynthesis;X Chen;Nat Mater,2022
2. Combining microtomography, 3D printing, and numerical simulations to study scale effects on the permeability of porous media;Luan C. de SM Ozelim;International Journal of Geomechanics
3. A 3D additive manufacturing approach for the validation of a numerical wall-scale model of catalytic particulate filters;Igor Belot;Chemical Engineering Journal
4. Ein Modell für das Gullstrandsche Normalauge [A model of Gullstrand’s "normal eye"];B. Rassow;Ophthalmologica,1972
5. 3D Printing: Print the future of ophthalmology;W Huang;Invest Ophthalmol Vis Sci,2014
Cited by
5 articles.