Abstract
Magnetic resonance imaging (MRI) is a powerful noninvasive diagnostic tool with superior soft tissue contrast. However, access to MRI is limited since current systems depend on homogeneous, high field strength main magnets (B0-fields), with strong switchable gradients which are expensive to install and maintain. In this work we propose a new approach to MRI where imaging is performed in an inhomogeneous field using radiofrequency spatial encoding, thereby eliminating the need for uniform B0-fields and conventional cylindrical gradient coils. The proposed technology uses an innovative data acquisition and reconstruction approach by integrating developments in field cycling, parallel imaging and non-Fourier based algebraic reconstruction. The scanner uses field cycling to image in an inhomogeneous B0-field; in this way magnetization is maximized during the high field polarization phase, and B0 inhomogeneity effects are minimized by using a low field during image acquisition. In addition to presenting the concept, this work provides experimental verification of a long-lived spin echo signal, spatially varying resolution, as well as both simulated and experimental 2D images. Our initial design creates an open MR system that can be installed in a patient examination table for body imaging (e.g., breast or liver) or built into a wall for weighted-spine imaging. The proposed system introduces a new class of inexpensive, open, silent MRIs that could be housed in doctor’s offices much like ultrasound is today, making MRI more widely accessible.
Publisher
Public Library of Science (PLoS)
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献