Abstract
Purpose
We recently constructed an algorithm to measure visual field (VF) using the variational Bayes linear regression (VBLR). This algorithm enabled a faster VF measurement than the Swedish interactive thresholding algorithm (SITA) standard while maintaining the test-retest reproducibility (Murata H, et al. Br J Ophthalmol 2021). The current study aimed to compare the structure-function relationship between the SITA standard and VBLR.
Method
In 78 eyes of 56 patients with primary open-angle glaucoma, VF measurements were conducted using both SITA standard and VBLR VF, as well as spectral-domain optical coherence tomography. The structure-function relationship was investigated between visual sensitivity and circumpapillary retinal nerve fiber layer in the whole VF. This analysis was repeated for each of the 12 sectors (30 degrees). The strength of the structure-function relationship was evaluated using the second-order bias-corrected Akaike Information Criterion (AICc) index.
Result
In the whole VF, AICc values of SITA standard and VBLR were 601.6 and 597.3, respectively. The relative likelihood that VBLR had a better structure-function relationship than the SITA standard was 88.2% (when the entire field was averaged) or 99.9% (when all test points were analyzed in the pointwise manner). With the sector-wise analysis, SITA standard had a better structure-function relationship than VBLR in 1 sector (Superior sector in the retina), whereas VBLR had a better structure-function relationship than SITA standard in 4 sectors (Supero-Nasal, Infero-Nasal, Inferior, and Infero-Temporal sectors) with >95% relative likelihood.
Conclusion
Although it depends on locations and similar between SITA standard and VBLR-VF, but VBLR-VF had a better structure-function relationship than the SITA standard overall.
Funder
the ministry of education, culture, sports, science and technology of Japan
the Ministry of Education, Culture, Sports, Science, and Technology of Japan
University of Tokyo
Japan Glaucoma Society
Publisher
Public Library of Science (PLoS)
Reference23 articles.
1. A new generation of algorithms for computerized threshold perimetry, SITA;B Bengtsson;Acta ophthalmologica Scandinavica,1997
2. SITA Fast, a new rapid perimetric threshold test. Description of methods and evaluation in patients with manifest and suspect glaucoma;B Bengtsson;Acta ophthalmologica Scandinavica,1998
3. Evaluation of a new perimetric threshold strategy, SITA, in patients with manifest and suspect glaucoma;B Bengtsson;Acta ophthalmologica Scandinavica,1998
4. Properties of perimetric threshold estimates from Full Threshold, SITA Standard, and SITA Fast strategies;PH Artes;Invest Ophthalmol Vis Sci,2002
5. Validating Variational Bayes Linear Regression Method With Multi-Central Datasets;H Murata;Invest Ophthalmol Vis Sci,2018