Author:
Dentel S. K.,Kingery K. M.
Abstract
In spite of the increased use of streaming current detectors (SCDs) as a means of monitoring and/or controlling coagulant dosage, knowledge regarding fundamental workings is incomplete. This paper provides an initial attempt at predicting and verifying functioning compared to electrophoretic mobility. The instrument's components -- the sensor and the signal processor -- are first described. Equations modelling electro-double layer behavior in its sensor are then developed. Simplifying assumptions include the use of a capacitance model of the double layer and a triangular velocity profile for fluid within the sensor's annulus. More complex modelling approaches are also suggested which incorporate the Gouy-Chapman electro-double layer model and an exact solution for the velocity profile. Experimental results confirm predictions of the simplified model under conditions of low potential. A monotonic relationship exists between streaming current electrophoretic mobility, which is required for its use as a control parameter. Deviations from model predictions are suggested to be due to charge characteristics of the sensor surfaces themselves.
Subject
Water Science and Technology,Environmental Engineering
Cited by
19 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献