Abstract
The ability to determine and control the normalized concentrations has been established using the experimentally obtained dependencies of the active and reactive constituents of conductivity in the wide frequency range of the electromagnetic field on the composition of the multicomponent fluid and the impurity concentrations. For high-aqueous substances dissolved in water (cooling liquids), a method for controlling the qualitative and quantitative composition is developed. It is based on the comparison of the measured and experimentally established active conductivity component and the comparison of the corresponding dependencies in the frequency field of the reactive component of conductivity. The developed electric method allows quantitative and qualitative estimation of the composition of the coolant on the contents of controlled components in the short time (up to 2 seconds) in a non-laboratory environment and to ensure uninterrupted operation of the equipment.
Publisher
IMEKO International Measurement Confederation
Subject
Electrical and Electronic Engineering,Mechanical Engineering,Instrumentation
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献