Rotational Resistance of Surface-Treated Mini-Implants

Author:

Kim Seong-Hun1,Lee Shin-Jae2,Cho Il-Sik3,Kim Seong-Kyun4,Kim Tae-Woo5

Affiliation:

1. a Postgraduate Student, Department of Orthodontics, School of Dentistry and Dental Research Institute, Seoul National University, Korea; Assistant Professor, Department of Orthodontics, The Catholic University of Korea, Uijongbu St. Mary's Hospital, Korea

2. b Associate Professor, Department of Orthodontics, School of Dentistry and Dental Research Institute, Seoul National University, Seoul, Korea

3. c Former Resident, Department of Orthodontics, School of Dentistry and Dental Research Institute, Seoul National University, Seoul, Korea

4. d Associate Professor, Department of Prosthodontics, School of Dentistry and Dental Research Institute, Seoul National University, Seoul, Korea

5. e Professor and Chair, Department of Orthodontics, School of Dentistry and Dental Research Institute, Seoul National University, Seoul, Korea

Abstract

Abstract Objective: To test the hypothesis that there is no difference in the stability and resistance to rotational moments of early loaded sandblasted and acid-etched (SLA) mini-implants and those of machined-surface implants of the same size and shape. Materials and Methods: A randomized complete block design was used in 12 skeletally mature male beagle dogs. Ninety-six orthodontic mini-implants were tested. Two types of implants were used: some had SLA surface treatment and some had machined surfaces without coating. After 3 weeks of healing, rotational moments of 150 g were applied. The success rates, maximum torque values, angular momentum, and total energy absorbed by the bone were compared. All values were subjected to mixed-model analysis to evaluate the influence of surface treatment, rotational force direction, and site of implantation. Results: The maximum insertion torque and angular momentum of SLA implants were significantly lower than those of machined implants (P = .034, P = .039). The SLA implants had a significantly higher value for total removal energy than the machined implants (P = .046). However, there were no significant differences in total insertion energy, maximum removal torque, and removal angular momentum between the 2 groups. There was no significant difference between clockwise and counterclockwise rotation in all measurements. Conclusion: SLA mini-implants showed relatively lower insertion torque value and angular momentum and higher total energy during removal than the machined implants, suggesting osseointegration of the SLA mini-implant after insertion.

Publisher

The Angle Orthodontist (EH Angle Education & Research Foundation)

Subject

Orthodontics

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3