Author:
Ab Shukor Nor Shazleen,Musarudin Marianie,Abdullah Reduan,Abdul Aziz Mohd Zahri
Abstract
This study aims to measure the radial dose function and anisotropy function F(r, θ) of high Dose Rate (HDR) 192Ir source in a fabricated water-equivalent phantom using Gafchromic® EBT3 film and TLD-100H and to compare the results obtained with the MCNP5 calculated values. The phantom was fabricated using Perspex PMMA material. For, the EBT3 films with a required dimension and TLD-100H chips were placed at r=1, 2, 3, 5, and 10 cm from the source. The F(r, θ) measurements were carried out at r=1, 2, 3, 5, and 10 cm with the angle range from 10° to 170°. The result of from EBT3 film and TLD-100H was in good agreement (2.10%±1.99). Compared to MCNP5, the differences are within 0.31% to 11.47% for EBT3 film and 0.08% to 10.58% for TLD-100H. For the F(r, θ), an average deviation with the MCNP5 calculation is 4.94%±2.7. For both and F(r, θ), the effects are prominent at r=10 cm. At this distance, the response of both Gafchromic® EBT3 film and TLD-100H shows less sensitivity as the dose followed the inverse square law. This work demonstrates that Gafchromic® EBT3 film dosimeter and TLD-100H are suitable dosimeters in 192Ir dosimetric measurements at a radial distance of ˂5 cm
Publisher
Universiti Putra Malaysia
Subject
General Earth and Planetary Sciences,General Environmental Science
Cited by
1 articles.