Optimisation of the drill-in behaviour of the EcoFit® SC threaded cup

Author:

Thorwächter Christoph1,Woiczinski Matthias1,Santos Inês1,Schmidutz Florian12,Paulus Alexander1,Jansson Volkmar1,Weber Patrick134

Affiliation:

1. Department of Orthopaedics, Physical Medicine and Rehabilitation , University Hospital, LMU Munich , Marchioninistr. 15 , 81377 Munich , Germany

2. BG Trauma Center Tübingen , University of Tübingen , Tübingen , Germany

3. ECOM-Excellent Center of Medicine , Munich, Arabellastr. 17 , 81925 Munich , Germany

4. ATOS Klinik Munich , Arabellastr. 17 , 81925 Munich , Germany

Abstract

Abstract Threaded cups show good clinical results when implanted correctly. In clinical use, multiple cases with an incomplete placement of the EcoFit threaded cup (implantcast) were observed. This behaviour could not be explained intra- and postoperatively. The aim of this study was to compare and optimise the drill-in-behaviour of the EcoFit cup in a biomechanical investigation. EcoFit cup sizes 46, 50 and 54 mm were compared with the SC cup (Aesculap) size 50 mm. Foam blocks (Sawbones) of density 0.16 g/ml (pcf 10), 0.32 g/ml (pcf 20) and 0.48 g/ml (pcf 30) were used. After standardised placement using a universal testing system (n = 8 per group), the primary stability, the overhang of the cups and the drill-in behaviour were measured. Overreamings of 1 and 2 mm were performed (pcf 20, n = 8) for the EcoFit cup size 50 and the primary stability as well as the overhang measurements were examined. Measurements of the cup diameter, thread depth and thread pitch were performed on three-dimensional (3D) images of the cup size 50 mm. The drill-in behaviour was different between the EcoFit and the SC cups. Even with maximum torque, the EcoFit cup could not be positioned as deep as the SC cup in standard reaming conditions (overhang of 1.1 ± 0.4 mm for the EcoFit size 50 in pcf 20 and of −0.01 ± 0.2 mm for the SC cup). The primary stability was lower for the EcoFit cup in comparison to the SC cup (128.8 ± 3.2 Nm vs. 138.6 ± 9.1 Nm, p = 0.0291). With overreaming to 51 mm, a deeper positioning of the EcoFit was possible (overlap of −0.3 ± 0.1, comparable to the SC cup). The overreaming of the cavity also led to a significantly higher primary stability of 143.4 ± 3.7 Nm (p < 0.001) comparable to the unaltered condition (128.8 ± 3.2 Nm). Overreaming to 52 mm had no further advantage in terms of primary stability or overhang. The geometric measurements showed significant differences as well. The previously clinically observed difficulties in inserting the cup were confirmed by this study. By overreaming to 51 mm, the drill-in behaviour, the primary stability and the measured overhang were comparable to the reference cup. The obtained results suggest that the extension of the acetabulum cavity to 51 mm while using the implantcast EcoFit size 50 should be implemented in clinical applications.

Publisher

Walter de Gruyter GmbH

Subject

Biomedical Engineering

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3