Evaluation of the analytical and clinical performance of a new high-sensitivity cardiac troponin I assay: hs-cTnI (CLIA) assay
Author:
Li Ling1, Shu Xin2, Zhang Litao2ORCID, Xu Ao1, Yang Juan1, Jing Yisha2, Wang Hui2, Zhang Zhenlu1
Affiliation:
1. Clinical Laboratory of Wuhan Asia Heart Hospital , Wuhan , Hubei , P.R. China 2. Clinical Laboratory of Wuhan Asia General Hospital , Wuhan , Hubei , P.R. China
Abstract
Abstract
Objectives
Cardiac troponin (cTn) is the key biomarker for diagnosis of acute coronary syndrome (ACS). We performed a complete assessment of the high-sensitivity cardiac troponin I (hs-cTnI) (CLIA) assay on the analytical performance and clinical diagnostic performance, which was compared with Abbott ARCHITECT hs-cTnI assay.
Methods
Sex-specific 99th percentile upper reference limits (URLs) were determined from a healthy population of 424 males and 408 females. High-sensitivity performance was assessed by examining the imprecision at sex-specific URLs and the detectable results above LoD in a cohort of healthy population. The diagnostic performance of the hs-cTnI (CLIA) assay was validated in a population of 934 patients with suspected ACS.
Results
The 99th percentile URLs were 15.3 ng/L for female, 31.3 ng/L for male and 24.2 ng/L for overall population. The total imprecision near the sex-specific 99th percentile URLs were <5 %. 76.74 % of females, 97.12 % of males and 86.69 % of overall population had cTnI values exceeding the LoD, which met the criteria of high-sensitivity troponin assay. No cross-reactivity or interference was identified. The diagnostic sensitivity, specificity, PPV, NPV, and AUC of hs-cTnI (CLIA) assay were 97.97 , 90.70, 79.02, 99.21 % and 0.9885, respectively, which were comparable to ARCHITECT hs-cTnI assay.
Conclusions
hs-cTnI (CLIA) assay is a high-sensitivity troponin I method with high precision, sensitivity and specificity. The clinical diagnostic performance of hs-cTnI (CLIA) is comparable to the established ARCHITECT hs-cTnI assay. Mindray’s hs-cTnI (CLIA) assay is an attractive alternative for diagnosis of myocardial infarction with a high level of accuracy and safety.
Funder
Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
Publisher
Walter de Gruyter GmbH
Subject
Biochemistry (medical),Clinical Biochemistry,General Medicine
Reference37 articles.
1. Fathil, MF, Md Arshad, MK, Gopinath, SC, Hashim, U, Adzhri, R, Ayub, RM, et al.. Diagnostics on acute myocardial infarction: cardiac troponin biomarkers. Biosens Bioelectron 2015;70:209–20. https://doi.org/10.1016/j.bios.2015.03.037. 2. Katrukha, IA. Human cardiac troponin complex. Structure and functions. Biochemistry 2013;78:1447–65. https://doi.org/10.1134/s0006297913130063. 3. Thygesen, K, Mair, J, Katus, H, Plebani, M, Venge, P, Collinson, P, et al.. Recommendations for the use of cardiac troponin measurement in acute cardiac care. Eur Heart J 2010;31:2197–204. https://doi.org/10.1093/eurheartj/ehq251. 4. Mair, J, Morandell, D, Genser, N, Lechleitner, P, Dienstl, F, Puschendorf, B. Equivalent early sensitivities of myoglobin, creatine kinase MB mass, creatine kinase isoform ratios, and cardiac troponins I and T for acute myocardial infarction. Clin Chem 1995;41:1266–72. https://doi.org/10.1093/clinchem/41.9.1266. 5. Mair, J, Genser, N, Morandell, D, Maier, J, Mair, P, Lechleitner, P, et al.. Cardiac troponin I in the diagnosis of myocardial injury and infarction. Clin Chim Acta 1996;245:19–38. https://doi.org/10.1016/0009-8981(95)06168-1.