Effects of calcium dobesilate (CaD) interference on serum creatinine measurements: a national External Quality Assessment (EQA)-based educational survey of drug-laboratory test interactions

Author:

Zhang Tianjiao1,Guo Xiuzhi2,Hou Li’an2,Zhao Haijian1,Ma Rong1,Xia Liangyu2ORCID,Li Honglei2,You Tingting2,Qiu Ling2,Zhang Chuanbao1

Affiliation:

1. National Center for Clinical Laboratories, Beijing Hospital, National Center of Gerontology , Institute of Geriatric Medicine, Chinese Academy of Medical Sciences , Beijing , PR China

2. Department of Laboratory Medicine , Peking Union Medical College Hospital, Peking Union Medical College, Chinese Academy of Medical Sciences , Beijing , PR China

Abstract

Abstract Objectives Drug-laboratory test interactions (DLTIs) are one of the major sources of laboratory errors. Calcium dobesilate (CaD) interference on serum creatinine testing is a widespread problem that has long been ignored in China. A national EQA-based survey was launched to investigate the current status of CaD interference on creatinine routine methods used in China and enhance the education of CaD interference in clinical laboratories. Methods A descriptive survey was developed to characterize the status quo of Chinese laboratory professionals’ cognition to CaD interference. Four of survey samples which were spiked with/without interference additive were shipped to 175 participant laboratories. The target reference values from a reference measurement procedure were compared against the results from participating laboratories to evaluate the CaD interference on serum creatinine measurements using enzymatic method or Jaffé method. Results The lack of knowledge of DLTIs and the barriers to collect information from pharmacological and laboratory data systems had become the main problems on implementing DLTIs education in China. A significant negative influence of CaD on enzymatic method was observed regardless of measurement platforms. Jaffé method was generally free from interaction with CaD but showed poor precision and accuracy at low creatinine concentrations. Conclusions More efforts should be made to enhance the education of DLTIs in clinical laboratories in China.

Funder

China Capital Health Research and Development of Special Project

Publisher

Walter de Gruyter GmbH

Subject

Biochemistry (medical),Clinical Biochemistry,General Medicine

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献