Falsely markedly elevated 25-hydroxyvitamin D in patients with monoclonal gammopathies

Author:

Hager Helle Borgstrøm1,Bolstad Nils2,Warren David J.2,Ness Marianne Vindal1,Seierstad Birgitte3,Lindberg Morten1

Affiliation:

1. Department of Medical Biochemistry , Vestfold Hospital Trust , Tønsberg , Norway

2. Department of Medical Biochemistry , Oslo University Hospital – Radiumhospitalet , Oslo , Norway

3. Department of Laboratory Diagnostics , Vestre Viken Hospital Trust , Drammen , Norway

Abstract

Abstract Objectives Monoclonal immunoglobulins can cause interference in many laboratory analyses. During a 4 month period we observed seven patients with monoclonal disease and falsely extremely elevated 25-hydroxyvitamin D (25(OH)D) results above 160 ng/mL (>400 nmol/L) measured using an immunoassay from Abbott Diagnostics. Based on these findings, we studied the occurrence of falsely elevated 25(OH)D in samples with paraproteins and investigated possible mechanisms of the observed interference. Methods 25(OH)D was analyzed using the Architect i2000 platform from Abbott Diagnostics and a higher order method, liquid chromatography-mass spectrometry (LC-MS/MS), in serum samples from 50 patients with known monoclonal disease. Patients with falsely elevated 25(OH)D were included in further studies to elucidate the cause of interference. Spuriously elevated results were in addition analyzed on two alternative platforms (Siemens and Roche). Results Falsely elevated 25(OH)D levels were present in eight patients on the Abbott analyzer and one on the Siemens platform. Results from Roche were comparable with LC-MS/MS. Additional investigations excluded elevated concentrations of rheumatoid factor and heterophilic antibodies as the cause of interference in the Abbott assay. Conclusions Laboratories should be aware of the risk of falsely elevated 25(OH)D in samples run on the Architect analyzer from patients with monoclonal disease. Highly elevated vitamin D results should be diluted and if the dilution is non-linear, rerun by a different method, preferably LC-MS/MS. In patients with spuriously elevated 25(OH)D without known monoclonal disease, the laboratory should consider requesting protein electrophoresis to exclude paraprotein interference.

Publisher

Walter de Gruyter GmbH

Subject

Biochemistry (medical),Clinical Biochemistry,General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3