Does a change in quality control results influence the sensitivity of an anti-HCV test?

Author:

Dimech Wayne J.1,Vincini Giuseppe A.2,Cabuang Liza M.2,Wieringa Megan3

Affiliation:

1. NRL, Australia , 4th Floor Healy Building, 41 Victoria Parade, Fitzroy , 3065 Melbourne, VIC , Australia , Phone: +61 3 9418 1132, Fax: +61 3 9418 1155

2. NRL, Australia , Melbourne, VIC , Australia

3. Monash Health, Monash Medical Centre , Melbourne, VIC , Australia

Abstract

Abstract Background Laboratories use quality control (QC) testing to monitor the extent of normal variation. Assay lot number changes contribute the greatest amount of variation in infectious disease serology testing. An unexpected change in six lots of an anti-HCV assay allowed the determination of the effect these lot changes made to the assay’s clinical sensitivity. Methods Two sets of seroconversion samples comprising of 44 individual samples and 9 external quality assessment scheme (EQAS) samples, all positive to anti-HCV, were tested in affected and unaffected assay lots, and the difference in the quantitative and qualitative results of the samples was analyzed. Results Of 44 low-positive seroconversion samples tested in affected and unaffected assay lots, only three samples had results reported below the assay cutoff when tested on two of the six affected assay lot. A further sample had results below the cutoff for only one affected lot. None of the EQAS samples reported false-negative results. Samples having a signal to cutoff value of less than 6.0 generally had lower results in the affected lots compared with the unaffected lots. Conclusions Unexpected changes in QC reactivity related to variation, in particular assay lot changes, may affect patient results. This study demonstrated that QConnect Limits facilitated the detection of an unexpectedly large variation in QC test results, allowed for the identification of the root cause of the change, and showed that the risk associated with the change was low but credible. The use of evidence-based QC program is essential to detect changes in test systems.

Publisher

Walter de Gruyter GmbH

Subject

Biochemistry (medical),Clinical Biochemistry,General Medicine

Cited by 7 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3