Affiliation:
1. Division of Laboratory Medicine , 68272 Hiroshima University Hospital , Hiroshima , Japan
2. Department of Clinical Practice and Support , 68272 Hiroshima University Hospital , Hiroshima , Japan
3. Application Support Division , Sysmex Corporation , Kobe , Japan
Abstract
Abstract
Objectives
Even in the current era of hematology analyzer automation and peripheral equipment, quality control sample measurement remains a manual task, leading to variability in quality control data and increased workload. In this study, we evaluated the performance of quality control measurement using the BT-50 Transportation Unit (BT-50, Sysmex, Kobe, Japan), equipped with a scheduled automatic quality control function, to ensure measurement accuracy and streamline the workflow of hematology testing.
Methods
We evaluated the automatic measurement performance of quality control samples using the BT-50 for six representative blood test parameters: WBC (white blood cell), RBC (red blood cell), HGB (hemoglobin), HCT (hematocrit), PLT (platelet), and RET% (reticulocyte percent). We evaluated the equivalence and compared measurement accuracy between the BT-50 and the manual method. We then compared the variability to other laboratories and confirmed the stability of quality control samples. We also evaluated changes in workflow and staff resources before and after the introduction of the BT-50.
Results
The quality control measurement results for the BT-50 and the manual method were found to be equivalent for all six parameters. The variability measured by the BT-50 was lower for some parameters compared to the manual method. Furthermore, the workflow was streamlined by reducing manual processes, resulting in increased efficiency.
Conclusions
We confirmed the performance of quality control measurements using the schedule function of the BT-50. Introducing the BT-50 reduced the operator’s workload, improved operational efficiency, and promoted the standardization of quality control measurements.