Analytical assessment of ortho clinical diagnostics high-sensitivity cardiac troponin I assay
Author:
Kavsak Peter A.12, Edge Tara3, Roy Chantele3, Malinowski Paul3, Bamford Karen3, Clark Lorna2, Lamers Shana3, Hill Stephen12, Worster Andrew1
Affiliation:
1. McMaster University , Hamilton , ON , Canada 2. Core Laboratory, Hamilton Health Sciences , Hamilton , ON , Canada 3. Clinical Research Laboratory and Biobank, Hamilton Health Sciences , Hamilton , ON , Canada
Abstract
Abstract
Objectives
To analytically evaluate Ortho Clinical Diagnostics VITROS high-sensitivity cardiac troponin I (hs-cTnI) assay in specific matrices with comparison to other hs-cTn assays.
Methods
The limit of detection (LoD), imprecision, interference and stability testing for both serum and lithium heparin (Li-Hep) plasma for the VITROS hs-cTnI assay was determined. We performed Passing-Bablok regression analyses between sample types for the VITROS hs-cTnI assay and compared them to the Abbott ARCHITECT, Beckman Access and the Siemens ADVIA Centaur hs-cTnI assays. We also performed Receiver-operating characteristic curve analyses with the area under the curve (AUC) determined in an emergency department (ED)-study population (n=131) for myocardial infarction (MI).
Results
The VITROS hs-cTnI LoD was 0.73 ng/L (serum) and 1.4 ng/L (Li-Hep). Stability up to five freeze-thaws was observed for the Ortho hs-cTnI assay, with the analyte stability at room temperature in serum superior to Li-Hep with gross hemolysis also affecting Li-Hep plasma hs-cTnI results. Comparison of Li-Hep to serum concentrations (n=202), yielded proportionally lower concentrations in plasma with the VITROS hs-cTnI assay (slope=0.85; 95% confidence interval [CI]:0.83–0.88). In serum, the VITROS hs-cTnI concentrations were proportionally lower compared to other hs-cTnI assays, with similar slopes observed between assays in samples frozen <−70 °C for 17 years (ED-study) or in 2020. In the ED-study, the VITROS hs-cTnI assay had an AUC of 0.974 (95%CI:0.929–0.994) for MI, similar to the AUCs of other hs-cTn assays.
Conclusions
Lack of standardization of hs-cTnI assays across manufacturers is evident. The VITROS hs-cTnI assay yields lower concentrations compared to other hs-cTnI assays. Important differences exist between Li-Hep plasma and serum, with evidence of stability and excellent clinical performance comparable to other hs-cTn assays.
Funder
Ortho Clinical Diagnostics
Publisher
Walter de Gruyter GmbH
Subject
Biochemistry, medical,Clinical Biochemistry,General Medicine
Reference34 articles.
1. Boeddinghaus, J, Twerenbold, R, Nestelberger, T, Koechlin, L, Wussler, D, Meier, M, et al. APACE investigators. Clinical use of a new high-sensitivity cardiac troponin I assay in patients with suspected myocardial infarction. Clin Chem 2019;65:1426–36. https://doi.org/10.1373/clinchem.2019.304725. 2. Boeddinghaus, J, Nestelberger, T, Koechlin, L, Wussler, D, Lopez-Ayala, P, Walter, JE, et al. APACE investigators. Early diagnosis of myocardial infarction with point-of-care high-sensitivity cardiac troponin I. J Am Coll Cardiol 2020;75:1111–24. https://doi.org/10.1016/j.jacc.2019.12.065. 3. Kavsak, PA, Mondoux, SE, Sherbino, J, Ma, J, Clayton, N, Hill, SA, et al. Clinical evaluation of ortho clinical diagnostics high-sensitivity cardiac troponin I assay in patients with symptoms suggestive of acute coronary syndrome. Clin Biochem 2020;80:48–51. https://doi.org/10.1016/j.clinbiochem.2020.04.003. 4. Wu, AHB, Christenson, RH, Greene, DN, Jaffe, AS, Kavsak, PA, Ordonez-Llanos, J, et al. Clinical laboratory practice recommendations for the use of cardiac troponin in acute coronary syndrome: expert opinion from the academy of the American association for clinical chemistry and the task force on clinical applications of cardiac bio-markers of the international federation of clinical chemistry and laboratory medicine. Clin Chem 2018;64:645–55. https://doi.org/10.1373/clinchem.2017.277186. 5. Krintus, M, Panteghini, M. Laboratory-related issues in the measurement of cardiac troponins with highly sensitive assays. Clin Chem Lab Med 2021;58:1773–83.