Abstract
Abstract
Background
Coronavirus disease 2019, abbreviated to COVID-19, represents an emerging health threat worldwide as, after initial reports in China, it has continued to spread rapidly. The clinical spectrum of the disease varies from mild to severe acute respiratory distress syndrome (ARDS). Moreover, many patients can be asymptomatic, thus increasing the uncertainty of the diagnostic work-up. Laboratory tests play a pivotal role in the diagnosis and management of COVID-19, the current gold standard being real-time reverse transcription polymerase chain reaction (rRT-PCR) on respiratory tract specimens. However, the diagnostic accuracy of rRT-PCR depends on many pre-analytical and analytical variables. The measurement of specific COVID-19 antibodies (both IgG and IgM) should serve as an additional, non-invasive tool for disease detection and management.
Methods
The imprecision of the MAGLUMI™ 2000 Plus 2019-nCov IgM and IgG assays (Snibe, Shenzhen, China) was assessed by adopting the Clinical and Laboratory Standards Institute (CLSI) EP15-A3 protocol. Linearity of dilution and recovery was evaluated by means of mixes of high-level pools and low-level pools of serum samples. Immunoglobulin time kinetics were evaluated using a series of serum samples, repeatedly collected from COVID-19-positive patients at different times, from <5 days up to 26–30 days.
Results
Findings at the analytical validation of the assay carried out according to the CLSI EP15-A3 guideline demonstrated that imprecision and repeatability were acceptable (repeatability was <4% and <6% for IgM and IgG, respectively, whilst intermediate imprecision was <6%). In addition, results of dilution and recovery studies were satisfactory. The kinetics of COVID-19 antibodies confirmed previously reported findings, showing a rapid increase of both IgM and IgG after 6–7 days from the symptom onset. IgG had 100% sensitivity on day 12, whilst 88% was the higher positive rate achieved for IgM after the same time interval.
Conclusions
The findings of this study demonstrate the validity of the MAGLUMI 2000 Plus CLIA assay for the measurement of specific IgM and IgG in sera of COVID-19 patients, and for obtaining valuable data on the kinetics of both (IgM and IgG) COVID-19 antibodies. These data represent a pre-requisite for the appropriate utilization of specific antibodies for the diagnosis and management of COVID-19 patients.
Subject
Biochemistry (medical),Clinical Biochemistry,General Medicine
Reference20 articles.
1. WHO. WHO Virtual press conference on COVID-19. March 11, 2020. https://www.who.int/docs/default-source/coronaviruse/transcripts/who-audio-emergencies-coronavirus-press-conference-full-and-final-11Mar2020.pdf?sfvrsn=cb432bb3_2 [accessed 16 Mar 2020].
2. Lippi G, Plebani M. The critical role of laboratory medicine during coronavirus disease 2019 (COVID-19) and other viral outbreaks. Clin Chem Lab Med 2020;58:1063–9.
3. Parodi SM, Liu VX. From containment to mitigation of COVID-19 in the US. J Am Med Assoc 2020. doi: https://doi.org/10.1001/jama.2020.3882. [Epub ahead of print].
4. Onder G, Rezza G, Brusaferro S. Case-fatality rate and characteristics of patients dying in relation to COVID-19 in Italy. J Am Med Assoc 2020. doi: https://doi.org/10.1001/jama.2020.4683. [Epub ahead of print].
5. Lauer SA, Grantz KH, Bi Q, Jones FK, Zheng Q, Meredith HR, et al. The incubation period of coronavirus disease 2019 (COVID-19) from publicly reported confirmed cases: estimation and application. Ann Intern Med 2020. doi: https://doi.org/10.7326/M20-0504. [Epub ahead of print].
Cited by
227 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献